The F-50 is a mid-size, water-cooled, three-phase compressor used primarily with Gifford-McMahon cryocoolers and MRI magnets Specification Cooling Method Water-cooled (requires 4–28°C inlet water) Electrical (Low Voltage - F-50L) 3-Phase 200V, 50/60 Hz Electrical (High Voltage - F-50H) 3-Phase 380/415V (50 Hz) or 460-480V (60 Hz) Typical Power Consumption 6.5–7.2 kW (50 Hz) / 7.5–8.3 kW (60 Hz) Helium Pressure (Static) ~1.60 - 1.65 MPa (at 20°C) Approx. 120 kg (264 lbs) 3. Maintenance Highlights
Loss of cooling capacity at the cryocooler cold head. Cause: Saturation or failure of the helium adsorber.
: Never tilt the compressor more than 30° during transport or installation to prevent internal oil migration and potential damage. F-50H Compressor Unit Technical Manual | PDF - Scribd
While original equipment manufacturer (OEM) PDF manuals are often restricted to authorized service distributors, this comprehensive technical guide provides the essential operational data, installation steps, and troubleshooting matrices found in the Sumitomo F-50 manual. 1. Technical Specifications & Requirements sumitomo f50 compressor manual free
: The unit compresses low-pressure helium gas returning from the cold head, cools it via a water-cooled heat exchanger, and removes oil mist through an internal oil separator and adsorber before returning high-pressure gas to the system.
This public link is valid for 7 days and shares a thread, including any personal information you added. This link or copies made by others cannot be deleted. If you share with third parties, their policies apply. Can’t copy the link right now. Try again later.
If you are looking for the , it is essential to understand the key technical details covered in the manual to maintain the unit properly. 1. Technical Specifications The F-50 is a mid-size, water-cooled, three-phase compressor
Proper installation is critical to prevent compressor overheating and oil contamination.
Always purge or pull a vacuum on lines before connecting to avoid introducing air or moisture into the helium loop. Routine Maintenance Procedures
: Delivers 6.5–7.2 kW at 50 Hz and 7.5–8.3 kW at 60 Hz. Cause: Saturation or failure of the helium adsorber
By providing this comprehensive guide, we hope to have helped you find the Sumitomo F50 compressor manual free or provided valuable information on this industrial equipment.
Spending $30 to avoid a single wrong part order (e.g., buying a $200 separator that doesn’t fit) is excellent value.